Blog

Author Archive

Reprocessed Single-Use Devices – Optimizing 510(k) preparation

This article explains the challenges reprocessors face in obtaining 510(k) clearance for reprocessed single-use devices when they are not the OEM.

Guidance for Reprocessed Single Use Devices Reprocessed Single Use Devices   Optimizing 510(k) preparation

With increasing pressures on the medical device industry to make healthcare more affordable, there has been a push to reprocess and reuse single-use devices. Reprocessors obtain used devices from healthcare facilities. The reprocessors clean, process, resterilize, repackage, and relabel devices. Reprocessors must obtain FDA 510(k) clearance by demonstrating that the safety and effectiveness of the reprocessed device are substantially equivalent to the single-use device produced by the original equipment manufacturer (OEM). The FDA also released a guidance document regarding reprocessed single-use devices.

Obtaining 510(k) clearance for a device your company did not design can be challenging because information requirements that are trivial for the OEM can be extremely difficult to provide for a reprocessor of the device. The following sections of a 510(k) submission pose unique challenges for reprocessed single-use devices:

Section 13, Labeling of reprocessed devices

Labeling of reprocessed devices consists of the instructions for use and the packaging label(s). Device package labeling may also direct the user to both the reprocessor’s IFU and the OEM’s IFU. If you are referencing the OEM’s IFU, it is also important to include the OEM’s model number. Instructions for use should consist of:

  1. Indications for use, which must be equivalent to the OEM indications.
  2. All of the necessary warnings and cautions and basic operating instructions needed to operate the device safely.
  3. The instructions for use may also instruct the user to reference the OEM instructions for use for additional information.
  4. Instructions on the handling of the device after use, with the likelihood that the device will be returned to the reprocessor to repeat the cycle.

Section 15, Biocompatibility

Biocompatibility data is more challenging to provide if you replace or modify original components. If reprocessing does not modify the OEM device whatsoever, you can claim that the materials are identical to the OEM device. Therefore, the reprocessed device does not require biocompatibility testing. However, the reprocessor still needs to evaluate the biological risks associated with the reprocessing of the device by testing for cleaning and sterilization residuals. This involves testing for cleaning agent residuals and EO residual testing (ISO 10993-7), if applicable.

If you replace any of the components during reprocessing, with a new component that is identical in dimension and material to the OEM component, minimal biocompatibility testing will be required. If the exact material used by the OEM is unknown, reprocessors can perform material identification testing to determine the material used, and then create the replacement part out of the same material.

If you modify or replace any patient-contacting components on the device such as lubricants, insulation, etc., with components that are different from the OEM, then you will need to perform additional biocompatibility testing to prove that the new or modified material is biocompatible. This testing will depend on the duration of contact and where will the material contact the patient. The new material will also need to be listed in your device description and Section 15 of your 510(k) submission.  

Section 18, Performance Testing

There are three primary sources for identifying performance testing requirements of reprocessed devices:

  1. OEM Testing listed in the OEM 510(k) submission
  2. Predicate Testing listed by another reprocessor of an equivalent device
  3. Product Standards listed under the product classification code for the reprocessed device or the OEM device

You should reference a predicate device that has been reprocessed and the OEM device to identify performance testing. Some testing is specific to the functional performance of the device. For these tests, you need to compare performance side-by-side against the OEM. Another testing is specific to the reprocessing, and you will reference the predicate device. Sources of information regarding the required tests for each of these devices can be found in the 510(k) summaries of the respective devices. If possible, it’s helpful to select a predicate that has a redacted 510(k) available on the FDA’s website. If a redacted 510(K) is not readily available, you may request a redacted copy through the freedom of information act on-line. A redacted copy of the OEM 510(k) is also helpful.

If testing information is not as readily available in the 510(k) summary, you will determine the essential performance functions of the device, and design tests to evaluate and compare the OEM device and the reprocessed device for those functionalities. Some devices have specific standards for the design and/or testing of the device. To determine if the reprocessed device has any applicable standards, you should search the product code of the reprocessed device, as well as the product code of the OEM device if they are different, in the FDA product classification database. Recognized standards applicable to the reprocessed device will be listed in the search results.

Additional tests that may be needed to validate reprocessing include residual protein, residual carbohydrates, and the presence of hemoglobin. These tests ensure that all biological material from previous use is removed. If you are not performing biocompatibility testing on the reprocessed device, you also need to do a chemical test to ensure no residual detergent or cleaning residues are remaining on the device. You also need to determine how many reprocessing cycles the device can survive before performance degradation. This can be done by repeating simulated use, reprocessing, and performance testing until a statistically relevant decrease in the performance of the device is observed.

If you have additional questions regarding the preparation of your 510(k) submission, you might be interested in a course Mary Vater and Rob Packard created for AAMI. Rob Packard will be the lead instructor for the course pilot in May: 510(k) training course. You can also schedule a call with us by clicking the button below.

Click here to schedule a 15 minute call 300x62 Reprocessed Single Use Devices   Optimizing 510(k) preparation

Posted in: 510(k)

Leave a Comment (0) →

Purchasing Controls and Supplier Qualification

This article identifies the requirements for purchasing controls and supplier qualification procedures, as well as best practices for implementation.

Purchasing Controls

Sourcing suppliers in the medical device industry is not as simple as going on the internet and finding your material and purchasing it. As part of a compliant quality management system, purchasing controls must be in place to ensure that quality products and materials are going into your device and that any service providers that your company uses in the production of your product or within your quality management system are qualified.

ISO 13485 Requirements

In light of that, ISO 13485:2016 sections 7.4.1 Purchasing process, 7.4.2 Purchasing information, and section 7.4.3 Verification of purchased product outline the purchasing requirements. The following are requirements for the evaluation and selection of suppliers:

  • The organization must have established criteria for the evaluation and selection of suppliers.
  • The criteria need to evaluate the supplier’s ability to provide a product that meets the requirements.
  • It needs to take into consideration the performance of the supplier.
  • It must consider the criticality and the effect that the purchased product may have on the quality of the medical device.
  • The level of supplier assessment and monitoring should be proportionate to the level of risk associated with the medical device.

Maintaining Purchasing Controls

To start, in the most basic sense, purchasing controls involve procedures that ensure you are only purchasing from suppliers who can meet your specifications and requirements. The best way to keep track of your qualified suppliers is to maintain an Approved Supplier List (ASL). You should only purchase products or services that affect your product or quality management system from companies on the ASL (you would not necessarily need to qualify things like office supplies or legal assistance through purchasing controls).

When used effectively, the Approved Supplier List can be a great tool to manage the key facets of purchasing control and keep track of supplier monitoring. Items that you can capture on the ASL include:

  • Supplier Name
  • Scope of Approved Supplies
  • Contact Information
  • Status of Approval (Approved, Pending, Unapproved, etc.)
  • Qualification Criteria
  • Supplier Certification and expiry dates
  • Monitoring Requirements/Activities
    • Date of Last Review
    • Date of Next Review

The first step in your purchasing procedure should involve checking to see if the supplier is under active approved status on the ASL. The second step will be to ensure that you are purchasing an item/service that is within the scope of approval of that supplier. If you have not approved the supplier, or the intended purchase is beyond the scope of that supplier, your purchaser will need to go through the necessary channels to add the supplier to the ASL or modify their scope on the ASL.

Supplier Qualification Criteria

As required by the FDA, the level of supplier assessment should be proportionate to the level of risk associated with the medical device. The FDA is not prescriptive about the use of specific qualifications or assessments for different types of suppliers, so that is up to your company to determine. This is a somewhat grey area but based on years working with companies and suppliers, as well as participating in FDA and ISO 13485 audits, there are some general expectations of vendor qualifications that we have observed and would recommend.

It is good practice to have a form or template that guides your supplier evaluation process. Using input from engineering and QA to first determine the level of risk and the requirements of that supplier, and then base your qualification plan on that information. If you have a higher risk supplier who may be supplying a critical component to your device, or providing a critical service such as sterilization, then your qualification process will be much more involved.

Here is an example of two different levels of criteria based on the type of supplier (the intent is not for the following items to be rules, and your company is responsible for determining the adequate acceptance criteria for suppliers, but this is a general example of what you may expect).

  • Critical Custom Component Supplier
    • ISO 13485 Certification
    • On-site audit of supplier’s facility
    • References
    • Provides Certificates of Analysis (CoA)
    • A written agreement that the supplier will communicate with the company regarding any changes that could affect their ability to meet requirements and specifications.
    • You validate a production sample, and it meets requirements
  • Non-Critical Consumable Supplier
    • Product available that meets the needs of the company.
    • An associate has previously used by an associate who recommends the supplier.
    • Adequate customer service returns allowed.

Additional Function of Supplier Evaluation Forms

The supplier evaluation form can also be used as the plan to assign responsibility and track completion and results during the initial evaluation. It can also include the plan for ongoing monitoring and control of the supplier. This evaluation form should be maintained as a quality record, and auditors will frequently ask to see supplier evaluations.

Are Supplier Audits Required as Purchasing Controls?

Also valuable, supplier audits may be included as part of an evaluation plan for a new supplier, the change of scope of a supplier, a routine audit as part of ongoing monitoring, or as part of a nonconformity investigation of a high-risk product. While it is not required by ISO 13485, nor does the FDA does specify in the CFR that you must audit suppliers, it is a very good idea to audit your critical suppliers. If an auditor or FDA inspector sees evidence that your current purchasing controls are inadequate, performing supplier audits may be forced as a corrective action.

Beyond that, you can gain so much value, and gather countless clues and important information in an audit that you just cannot get without visiting your critical supplier. You can see where they plan to/are making/cleaning/sterilizing/storing your product. Talk to the people on the line, are they competent and trained? Does the company maintain their facility well? How secure is it? Do they maintain adequate records and traceability? Have there been any nonconformities relating to your product that have been detected? Etc.

Supplier audits should also include evaluation of the procedures, activities, and records of the supplier that could have an impact on the product or service they are providing your company. If it is not the first audit of the company, you should be sure to review the previous audit report findings and ensure the company has addressed any nonconformities, review supplier performance data, information about any changes that may have occurred at the supplier since your last visit, etc.

Record Maintenance and Ongoing Evaluation of Suppliers

No matter the method of supplier qualification, it is best practice to maintain supplier files that contain useful information relative to the supplier that may include:

  • The original supplier qualification form
  • Supplier certificates
  • References
  • Audit reports
  • Subsequent performance evaluations
  • Expanded scope qualifications
  • Supplier communications
  • Current contact information
  • Copies of any non-conforming material reports related to the supplier, etc.

ISO 13485 requires monitoring and re-evaluation of suppliers, and maintaining detailed supplier files will assist in meeting this requirement, and will help in the feedback system to identify and recurring problems or issues with a supplier. On a planned basis, whether that is annually, or every order (dependent on the criticality of the product), your company should conduct a formal supplier evaluation to determine whether the supplier has continued to meet requirements – In general, annual supplier reviews are standard. Additionally, you must specify this frequency in your procedure (auditors will look for what period you specify in your procedure, and then will check your ASL to make sure all of your suppliers have been reviewed within that timeframe).

During the supplier evaluation, if you find there have been issues, you need to determine and weigh the risks associated with staying with that supplier, and document that in the supplier file. If you determine the supplier should no longer be qualified, then you must also indicate on the ASL that the company no longer approves of the supplier.

Making the Purchase

When you have verified your supplier is approved on the ASL, you are authorized to purchase a product. Engineering is usually responsible for identifying the product specifications, requirements for product acceptance, and adequacy of specified purchasing requirements before communication to the supplier. The specifications may be in the form of drawings or written specifications. Additional information communicated to the supplier should also include, as applicable, an agreement between your company and the supplier that the supplier will notify you before the implementation of changes relating to the product that could affect its ability to meet specified purchasing requirements. When the first batch of product is received from a particular supplier, it is a good idea to verify that the product performs as intended before entering into production with new material or components.

Supplier Nonconformity

From time to time, you may encounter issues with a supplier. Sources of nonconformity include incoming inspections, production nonconformities, final inspection, or customer complaints. You must notify your supplier of the nonconformity and record their response and assessment. Depending on the level of criticality of the vendor, it is reasonable to require them to perform a root cause analysis to determine and alleviate the cause of failure. You should also request documentation of an effectiveness check to ensure the supplier has taken corrective actions. You should maintain copies of supplier nonconformity reports in the supplier file, and discuss nonconformities during ongoing supplier evaluations.

If the supplier does not cooperate or fails to address the nonconformity in an acceptable manner, or if there is a pattern of nonconformities with the vendor, then you should disqualify the supplier, and indicate that the supplier is “not approved” on the ASL.

Purchasing Controls Procedures You Might Need

Medical Device Academy developed a Supplier Qualification Procedure, Purchasing Procedure, and associated forms that will meet purchasing controls regulatory requirements for ISO 13485:2016 and 21 CFR 820.50. These procedures will help you ensure that goods and services purchased by your company meet your requirements and specifications. If you have any questions or would like help in developing a custom procedure or work instructions that meet your company’s unique needs, please feel free to email me or schedule a call to discuss.

Posted in: ISO 13485:2016

Leave a Comment (1) →

Risk Management Requirements – 510k vs DHF

This article compares the risk management requirements for a 510k submission with the risk management requirements for your design history file (DHF).

Design Controls and Risk Management Risk Management Requirements   510k vs DHF

Risk Management Requirements and Design Control Requirements

Last week I presented a free webinar on how to combine risk management with design controls when planning to submit a 510k. Many questions were asking what the design control and risk management requirements are for a 510k.

What are the Design Control Requirements in a 510k?

There is no specific part of the regulations stating what the 510k design control requirements are. However, some aspects of the DHF are required as 510k design control documentation, but not necessarily in the exact form as maintained in the DHF. For example, Design Inputs and Design Outputs are presented as applicable recognized standards and design specifications, while others will remain precisely the same (i.e., verification and validation test reports).

What are the Risk Management Requirements in a 510k?

For 510k submissions, the only risk management requirements are the inclusion of risk documentation for devices containing software of at least moderate level risk. There are some exceptions to this as well, though, based on a few special control guidance documents—especially when the submission type is an abbreviated 510k. This is article identifies which of the DHF and RMF elements are 510k design control requirements and 510k risk management requirements.

510k Design Control Requirements

Design Controls are identified in 21 CFR 820.30. Every manufacturer of any class II or class III devices and certain Class I devices (class I devices with software, tracheobronchial suction catheters, surgeon gloves, protective restraints, radionuclide applicators, radionuclide teletherapy devices) need to control design per this regulation. The requirement for a Design History File is item j) and states:

“Each manufacturer shall establish and maintain a DHF for each type of device. The DHF shall contain or reference the records necessary to demonstrate that the design was developed following the approved design plan and the requirements of this part.”

The “requirements of this part” refer to the other bullets in 21 CFR 820.30 which can be summarized as:

a) Establish and maintain procedures to control the design of a device.

b) Design and Development Planning – Each manufacturer shall establish a plan that describes the design and development activities, and defines responsibilities for implementation.

c) Design Inputs – Manufacturers need to ensure design requirements relating to a device are appropriate and address the intended use of the device.

d) Design Outputs – Design outputs need to be documented in terms that allow an adequate evaluation of conformance to design input requirements. Design outputs that are essential for the proper functioning of the device should be identified.

e) Design Review – Formal documented reviews of design results should be planned and conducted at appropriate stages of device development.

f) Design Verification – Design verification confirms that the design output meets the design input requirements.

g) Design Validation – Design validation shall be performed under defined operating conditions on initial production units or their equivalents. It shall ensure that devices conform to defined user needs and meet the intended use of the device.

h) Design Transfer – Design transfer documentation shall ensure that the device design is correctly translated into production specifications.

i) Design Changes – changes should be identified, documented, validated/verified, reviewed, and approved before their implementation.

The Design History File is intended to be a repository of the records required to demonstrate compliance with your design plan and design control procedures. While companies are required to create and maintain this documentation according to the FDA regulation, not all of the documentation will be reviewed as part of the 510k. The following table compares the elements that comprise a DHF with the 510k design control requirements.

DHF Element 510k Design Control Requirements
Design Plan Not Required
User Needs Not Required
Design Inputs

Cover Sheet (Section 1) and

Declaration of Conformity (Section 9)

Some design inputs will appear in the form of standards in FDA Form 3514 (Cover Sheet) and the Declaration of Conformity FDA Form 3654 (Standards Data Report)

Design Outputs

Device Description (Section 11)

The Device Description lists the specifications of the device, and your Design Outputs document will help populate the Device Description. This can include drawings, pictures, or written specifications that describe your device.

Labeling

Proposed Labeling (Section 13)

The labeling is usually considered part of the Design Outputs within the DHF and is included specifically in the labeling section of the 510(k) submission. This includes both the Instructions for Use and any Package Labeling.

Verification and Validation Protocols

Not Required

You do not have to include the protocols, but the reviewer may ask to see them if they have any questions when reviewing the reports.

Verification and Validation Reports

Sterilization (Section 14)

Biocompatibility (Section 15)

Software (Section 16)

Electrical Safety and EMC (Section 17)

Bench Performance Testing (Section 18)

Animal Performance Testing (Section 19)

Clinical Performance Testing (Section 20)

Of course, not all of these sections will be applicable to every device. Still, you should include all relevant validation test reports within your submission in the appropriate part of the 510k. Typically, each of these sections will have a cover sheet that outlines the reports that are included within the section, and then you can just include the report from the DHF in its entirety behind the cover sheet in that section.

Process Validation Only required for sterilization validation typically, but there are exceptions for novel materials and coatings
Work Instructions Not Required for 510k
Design Review Meeting Minutes Not Required for 510k
Design Trace Matrix Only required for software
Risk Management File Sometimes – See Risk Management File Table Below
Post-Market Surveillance Plan Not Required, but a few exceptions for high-risk devices
Clinical Data Summary Required only if used to demonstrate safety and efficacy
Regulatory Approval It Will result from 510k Clearance, so nothing to be included in 510k submission.

510k Risk Management Requirements

Regarding the FDA regulations for risk management, there is a requirement under the Design Validation section of 21 CFR 820.30 that states:

“Design validation shall include software validation and risk analysis, where appropriate.”

For FDA compliance and CE Marking, both recognize ISO 14971 as the standard for risk management. FDA recognizes ISO 14971:2007 whereas EN ISO 14971:2012 is the European National version for CE Marking. Rob Packard wrote an article describing the contents of the risk management file as well as the specific differences in the requirements between the FDA and CE Marking with regard to ISO 14971.

For your 510k submission, the FDA only requires risk management documentation to be included if the product contains software, and the risk is at least a level of “moderate concern”. There are some other cases when risk management is required by special controls guidance documents, but even when it is required, you only have to submit your risk analysis. The table below describes the risk management requirements in greater detail.

RMF Element 510k Risk Management Requirement
Risk Management Plan Not Required
Hazard Identification

510ks with Software Only (Section 16)

Hazard Identification is only required for devices that have a software component. It is not required for most other devices.

Risk Assessment

510(k)s with Software (Section 16)

Certain Special Controls Guidance

The Risk Assessment is only required to be included in your device contains software, or if a special controls guidance document specifically requires a risk assessment. It is not required for other 510ks.

Risk Control Option Analysis Software and Certain Special Controls Guidance
Risk Control Verification and Validation

Sterilization (Section 14)

Biocompatibility (Section 15)

Software (Section 16)

Electrical Safety and EMC (Section 17)

Bench Performance Testing (Section 18)

Animal Performance Testing (Section 19)

Clinical Performance Testing (Section 20)

This will not be any additional or special documentation specific to Risk Management and was already included in the DHF breakdown above. Still, the verification and validation also relate to risk management in ensuring that the risks have been adequately mitigated.

Risk-Benefit Analysis

Not Required for 510(k)

Risk-Benefit analyses are only required for De Novo applications, Humanitarian Device Exemptions, and PMAs.

Informing Users and Patients of Risks

Labeling (Section 13)

Part of the risk management will appear in the Labeling section of the 510k as warnings, contraindications, and precautions within the Instructions for Use and Package Labeling.

Risk Management Report Not Required

Special Controls Guidance Documents with Risk Management Requirements

Your first step in preparing your 510k submission is to search the FDA Guidance Document Database to determine if there is an applicable guidance document for your device. You can read another blog we wrote to explain Special Controls Guidance documents, and how to determine if one applies to your device. The following list provides examples of Class II Special Controls Guidance documents that require risk analysis to be included within the 510k:

When there are 510k risk management requirements, the special controls guidance document will typically state, “We recommend that the summary report contain:

An identification of the Risk Analysis method(s) used to assess the risk profile in general as well as the specific device’s design and the results of this analysis. (Refer to Section 6 for the risks to health generally associated with the use of this device that FDA has identified.)

Discussion of the device characteristics that address the risks identified in this class II special controls guidance document, as well as any additional risks identified in your risk analysis.”

The special controls guidance will also identify risks to health that have been identified for products of that type, which you should be sure to include in your risk analysis as appropriate.

More Information on Design Control and Risk Management Requirements

Hopefully, you are now able to determine which elements of your DHF are 510k design control requirements and which elements of your RMF are 510k risk management requirements. If you would like more information about how to implement design controls and risk management within your product development process, please consider registering for one of our training webinars:

If you need any further information or specific assistance with your 510k submission, please feel free to send me an email at mary@fdaecopy.com or schedule a call with our principal consultant, Rob Packard. He can answer any of your medical device regulatory questions.


Click here to schedule a 15 minute call 300x62 Risk Management Requirements   510k vs DHF

Posted in: 510(k), Design Control, ISO 14971:2019 (Risk Management)

Leave a Comment (0) →

Cybersecurity FDA Guidance for Devices with Software and Firmware

This article reviews the 2014 FDA guidance for premarket and post-market cybersecurity of medical devices with software and firmware—including requirements for reporting field corrections and removals.

Cybersecurity with custom aspect ratio Cybersecurity FDA Guidance for Devices with Software and Firmware

Hospitals, home health systems, and medical devices are more connected now than ever. The automatic communication between medical devices and network systems is improving efficiency and accuracy in the world of healthcare. Medical devices are capable of more computing, analysis, reporting, and automation to improve the speed and quality of patient care. There are even devices that consist only of software (i.e. software as a medical device or SaMD). Along with technological advances, new risks and concerns are also introduced. The risk of hackers exploiting vulnerabilities in networks and software is inevitable. The FDA introduced guidance for both pre-market and post-market cybersecurity to assist manufacturers in developing effective controls to protect patients and users. Cybersecurity protection requires Identification, Protection, Detection, Response, and Recovery.

The first step is incorporating processes and procedures to improve device cybersecurity into your quality management system. You should have a specific cybersecurity plan (i.e. security risk management plan) to outline the steps necessary to ensure a safe and secure medical device. In addition, your software development team will need cybersecurity training. The only medical device guidance document specific to cybersecurity is currently AAMI TIR57:2016.

Identify Cybersecurity Risks

The key to understanding and assessing the cybersecurity risks involved with your device begin in the early stages of design development. At the start of the risk management process, you need to identify the essential safety and performance requirements of the device. You need to identify any potential cybersecurity vulnerabilities that could impact safety or performance, as well as the specific harms that could result if the vulnerability was exploited. In assessing the specific vulnerabilities, the FDA recommends using the Common Vulnerability Scoring System (CVSS). There is a CVSS calculator available online through NIST. The overall score is calculated based on different factors such as attack vector (local, adjacent network, network), access complexity (high, medium, low), authentication (multiple, single, none), the impact of confidentiality (none, partial, complete), exploitability (unproven that exploit exists, proof of concept code, functional exploit exists), remediation level (official fix, temporary fix, workaround, unavailable), collateral damage potential (low, medium, high), etc. This score is used in the hazard analysis in determining the level of risk.

Cybersecurity Protection

The process of assessing the exploitability and harms can also assist in determining mitigations that can be implemented to reduce the cybersecurity risk. During the design process, the FDA expects you to implement as many protections as practicable. Protections include:

  • Limit Access to Trusted Users
    • Password protection strengthened password requirements
    • User authentication
    • Layered privileges based on user role
  • Limit Access to Tampering
    • Physical locks on devices and/or communication ports
    • Automatic timed methods to terminate sessions
  • Ensure Trusted Content
    • Restrict software or firmware updates to authenticated code
    • Systematic procedures for authorized users to download software and firmware only from the manufacturer
    • Ensure capability of secure data transfer, use of encryption

Cybersecurity Detection

The FDA also requires you to implement features that allow for security compromises to be detected, recognized, logged, timed, and acted upon during regular use. You should develop and provide information to the end-user concerning appropriate actions to take upon the detection of a cybersecurity event. Methods for retention and recovery should be provided to allow recovery of device configuration by an authenticated privileged user.

If you include off-the-shelf (OTS) software in your device, you are responsible for the performance of the software as part of the device. All software changes to address cybersecurity vulnerabilities of the OTS software need to be validated. You need to maintain a formal business relationship with the OTS vendor to ensure timely notification of any information concerning quality problems or corrective actions. Sometimes you will need to involve the OTS vendor to correct cybersecurity vulnerabilities.

Post-Market Surveillance

Once you complete the hazard analysis, mitigation implementation, validations, and has deployed their device for use – your activities shift to post-market management. Several QMS tools can assist in the cybersecurity processes post-market, including complaint handling, quality audits, corrective and preventive action, ongoing risk analysis, and servicing. A critical component of every cybersecurity program is the monitoring of cybersecurity information sources to assist in the identification and detection of risk. You should maintain contact with third-party software suppliers for the identification of new vulnerabilities, updates, and patches that come available.

There are many sources that companies should follow for information relating to cybersecurity, including independent security researchers, in-house testing, software or hardware suppliers, healthcare facilities, and Information Sharing and Analysis Organizations (ISAO). Involvement in ISAOs is strongly recommended by the FDA and reduces your reporting burden if an upgrade or patch is required post-market. ISAOs share vulnerabilities and threats that impact medical devices with their members. They share and disseminate cybersecurity information and intelligence pertaining to vulnerabilities and threats spanning many technology sectors, and are seen as an integral part of your post-market cybersecurity surveillance program.

Response and Recovery

If you identify a cybersecurity vulnerability, there are remediation and reporting steps that need to occur. Remediation may involve a software update, bug fixes, patches, “defense-in-depth” strategies to remove malware, or covering an access port to reduce the vulnerability. Uncontrolled risks should be remediated as soon as possible and must be reported to the FDA according to 21 CFR 806. Certain circumstances remove the reporting requirement. The decision flowchart below can be used to determine the reporting requirements.

Cybersecurity software change decision tree Cybersecurity FDA Guidance for Devices with Software and Firmware

In addition to reporting corrections and removals, the FDA identifies specific content to be included in PMA periodic reports regarding vulnerabilities and risks. If you have a Class III device, you should review that section thoroughly to ensure annual report compliance.

If a device contains software or firmware, cybersecurity will be an important component of the risk management processes, and continual cybersecurity management will be necessary to ensure the ongoing safety and effectiveness of your device. If you need more help with cybersecurity risk management of your medical device, please schedule a free 30-minute call with Medical Device Academy by clicking on the link below.

Click here to schedule a 15 minute call 300x62 Cybersecurity FDA Guidance for Devices with Software and Firmware

Posted in: Software Verification and Validation

Leave a Comment (0) →