CE marking

EN standard – What is it?

CE Marking auditors may ask if you assessed the difference between the ISO version of a standard and the EN standard. Is there a difference?

EN ISO 14971

What is an EN standard?

European Standards are technical standards ratified by one of the three European standards organizations: CEN, CENELEC, or ETSI. European Standards are referred to as an “EN standard.” The “EN” is derived from the German name Europäische Norm (i.e., translated as “European Norm”). Each of the member states have their own standards organizations that is responsible for adopting ISO standards that have been ratified, translation of the standard into the language of the member state, and and issue of translated EN standard. For example: German standards are preceded by “DIN,” Irish standards are preceded by “I.S. EN,” and Swedish standards are preceded by “SS-EN.”

Click here to subscribe to our blog feed

How is an EN version different from the ISO version of a standard?

Historically, EN medical device standards include three Annexes related to harmonization with the standard with the three EU Directives (i.e., MDD, AIMD, and IVDD). Now that the EU Device Regulations have been released (i.e., MDR = 2017/745) and (IVDR = 2017/746), the EN standards now have harmonization Annexes for the two regulations (i.e., ZA and ZB). The content of the ISO Standard is usually not changed in an EN standard unless there is a correction, amendment, or deviation.

Where can you purchase EN versions?

EN standards are translated from the adopted ISO standard by the standards organization for each member state. Below are a few examples:

If you already own the ISO version, do you need to buy the EN standard too?

In a Technical File for CE Marking, you are required to demonstrate how you comply with Annex I of the MDR or IVDR. The recommended method of demonstrating compliance is creating a General Safety and Performance Requirements (GSPRs) checklist. In that checklist, you need to identify which applicable standards were used. If an EN standard is available, the GSPR checklist should reference that standard instead of the ISO version. Unfortunately, in order to claim compliance with the EN version, someone needs access to that standard. This could be within your organization or a consultant that assisted in preparing the GSPR checklist to ensure that you are compliant with the EN standard. Generally, we purchase English versions of EN standards from the Estonian Center for Standards and Accreditation, because it is usually the least expensive source. However, if you ask a consultant to do this comparison for you, the best way to perform that comparison is using the comparison function of Adobe Acrobat Pro.

Can you identify EN requirements without the EN version?

Most of the quality system requirements for the MDR and IVDR regulations are found in Article 10 of the regulation. However, there are quality system requirements found in other articles and in the annexes. Therefore, you may be able to find EU requirements in the regulations by doing a keyword search. For example, searching for the word “risk” may help you find risk management requirements throughout the regulations. You may also find European-specific requirements in Common Specifications and MDCG guidance documents.

How to respond to a certification body auditor?

Final Answer: I’m not sure, because every auditor is a little different in their approach. However, as an instructor, I would teach an auditor to ask open-ended questions, such as: “How did you determine if there is an impact upon your procedures and design documentation with regard to the EN standard?” (i.e., – impact analysis). If the company provides an impact analysis and explains why the existing documentation and procedure should not change, I believe this meets the EU requirements. If the certification body auditor is still not satisfied, then you might try asking them, “What differences are you aware of between the EN and ISO versions of the standard?”

 

EN standard – What is it? Read More »

How Declaration of Conformity (DoC) templates are created

This article about the declaration of conformity for medical device CE Marking identifies three possible sources for creating a template and includes recommendations for making the transition to the proposed EU Medical Device Regulation (EMDR).

declaration of conformity for medical devicesCurrent declaration of conformity requirements

Each of the three “New Approach” Device Directives requires that manufacturers create a Declaration of Conformity (DoC). For the Medical Device Directive (MDD), this requirement is found in Annex II, IV, V, VI, and VII. The “Declaration of Conformity” is part of the title for each of those Annexes. A DoC is required for each CE Marked product family, but the MDD does not guide the format or content. Manufacturers need a controlled template.

Click here to subscribe to our blog feed

Three sources of declaration of conformity examples

There are three possible sources for creating a declaration of conformity template:

  1. Copy another company’s declaration of conformity
  2. Authorized Representatives
  3. Annex III of the proposed European Medical Device Regulations

Most companies are using option number 1—often without knowing it. The problem with this approach is you have no regulatory reference to justify the basis for the content and format of your Declaration.

The second option is to ask your Authorized Representative (AR). The AR is supposed to verify that you have a DoC for each product family and that the company has the required technical documentation for each product family. The AR must provide a DoC for each product they represent to the Competent Authority upon request. Therefore, some ARs provide manufacturers with a template for the DoC to ensure that your DoC meets requirements. This approach gives you a reference of external origin, and you can be sure that the template will meet current requirements.

My new recommendation for Creating A declaration of conformity

Option 3 is my new recommendation for companies. The proposed European Medical Device Regulation (EMDR) was released on September 26, 2012. The proposal provides a prescriptive template for the content of a DoC. This will eventually be required for every DoC. Annex III is only one page long, and there are only ten requirements:

  1. Name of the Manufacturer and the AR
  2. A declaration that DoC Issued under Sole Responsibility of Manufacturer
  3. Unique Device Identifier (UDI)
  4. Product Name and Catalog Numbers
  5. Risk Classification
  6. Declaration Statement of Conformity
  7. Reference to Harmonized Standards and Common Technical Specifications (CTS)
  8. Notified Body Name and Number, Description of Conformity Assessment Procedure and Certificate
  9. Additional Information
  10. Place and Date of Issue, along with the Name and Function of the Person Signing

Instead of waiting for the final approval of the EMDR, I recommend creating a declaration of conformity (DoC) template now that matches the proposed Annex III (you might want to add this to recommendations for improvement in your next management review).

Brigid Glass is the subject matter expert behind most of the conventions that our consulting firm uses for document templates. As with all of our forms, we assign a document control number (e.g., FRM-001), and the revision is “D1” for the first draft. The template is in table format to facilitate easier review by auditors and your AR. The left column of the table lists each of the required elements, and the right column is where the variable data is entered. Variable fields that are examples are in green font, and the content that does not require changing is in black font. There are some comments to explain how to fill in the variable content, and several items that will change when the proposed regulation is approved are identified. The following harmonized symbols from ISO 15233-1 are used in the template, as well:

DoC

Please click here to download the Medical Device Academy’s declaration of conformity (DoC) template.

 

 

 

 

How Declaration of Conformity (DoC) templates are created Read More »

Risk Control Selection – Deviation #5 in ISO 14971

ISO 14971:2012 deviation #5 is specific to selecting risk control options and protective measures for CE Marking medical devices.

Deviation #5 is regarding risk control option analysisIf your company is CE Marking medical devices, you are required to satisfy the Essential Requirements for Safety and Performance as defined in the three European Directives: the MDD, the AIMD, and the IVDD. Throughout these Essential Requirements, there is a requirement to reduce risks “as far as possible” (AFAP) by implementing risk controls. At one time, the expectation was for companies to implement state of the art concerning risk controls, and “state of the art” was interpreted as the latest version of the harmonized ISO Standards. However, lawyers dominating the European Commission appear to disagree with the status quo.

Click here to subscribe to our blog feed

Therefore, in 2012, the European National (EN) version of the Medical Device Risk Management Standard was revised. There is no change to the content of Clauses 1 through 9. Instead, the European Commission identified seven content deviations between the ISO 14971 Standard and the EU Directives. These deviations are identified and explained in Annexes ZA, ZB, and ZC. This blog is the fifth installment of Medical Device Academy’s seven-part blog series on this topic. The goal of the series is to identify solutions for meeting the Essential Requirements by suggesting changes to the current best practices of implementing a risk management process for medical device design.

Discretion as to the Risk Control Options/Measures

Essential Requirements 1 and 2 require that risk control options are implemented for all risks before determining the acceptability of residual risks. The 2nd Essential requirement also requires manufacturers to implement all risk control options—unless the risk controls do not further reduce risk.even if risks are acceptable you must still perform risk control option analysis

Clause 6.2 of the 14971 Standard suggests that you only need to use “one or more” of the risk control options, and Clause 6.4 indicates that further risk control measures are not required if the risk is acceptable. There is an apparent contradiction between the intent of the Standard and the Directives.

If risk acceptability has no impact on whether you will implement risk controls, there is no need for performing a preliminary risk evaluation. Therefore, I have three recommendations for changes to your current risk management process:

  1. Ignore Clause 5 of the 2007/2009 version of ISO 14971
  2. Eliminate the second step of risk assessment from your flow chart for risk management (see Figure 1 from the 14971 Standard)
  3. Define risk management policies upon clinical benefits, rather than absolute risks

Instead of performing a preliminary risk evaluation (Clause 6.5), risk/benefit analysis should be moved to Clause 7, where the evaluation of overall residual risk acceptability is required. By making this change, risk controls will be implemented, regardless of risk acceptability, and the acceptability of risks will be dependent upon the risk/benefit analysis alone.

Impact of this Deviation

Implementing changes to your risk management process to address this deviation has great potential to impact the design of devices—not just the risk management documentation. Design teams will no longer be able to stop the design process with an initial design that has an “acceptable risk.” Instead, design teams will be forced to implement additional risk controls and protective measures for device designs that already have a low risk of harm for specific failure modes.

The requirement to implement additional risk controls will increase the cost of devices that may have been relatively safe without the risk controls. For example, if a device is not intended to be implanted, it is a potential foreseeable misuse. Your company may have used the instructions for use to communicate the residual risk associated with misuse of the device. However, now your company will have to implement design controls (e.g., –a selection of materials suitable for implantation) to eliminate the risks associated with misuse and protective measures (e.g., – radio-opaque thread) to help retrieve product that was implanted in an “off-label” usage.

If you are interested in risk management training, Medical Device Academy offers a risk management training webinar.

Risk Control Selection – Deviation #5 in ISO 14971 Read More »

Which Countries Require CE Marking of Medical Devices?

 

28 Member States 2013

This blog serves as a reference guide with a discussion of information resources and a list of which countries require CE marking of medical devices.

Click here to subscribe to our blog feed

You can locate the current list of countries requiring CE Marking of medical devices by visiting the list of Competent Authorities (CAs) on the Europa webpage (http://bit.ly/ContactPoints). That page has 33 national CAs identified. CAs are the US FDA equivalent in the European Union (EU). In addition to member states in the EU, the CAs list also includes signatories (i.e., – countries that have signed the 1985 Schengen Agreement to allow people to pass between countries with no border controls) and EU candidate member states. For the most current status of candidate member states and potential candidate member states, you can visit the following Europa webpage: (https://european-union.europa.eu/principles-countries-history_en). As of September 21, 2013, the status of the 33 CAs is categorized in the list at the end of this blog posting.

Australia-EU Mutual Recognition Agreement

In addition to the 33 countries listed below, the Australia Therapeutic Goods Administration (TGA) has a mutual recognition agreement with the EU—the EC MRA (http://bit.ly/TGA-EU-MRA). This agreement, however, has limitations. The deal includes a rule of origin clause that excludes products manufactured outside the EU and Australia. Other restrictions include:

  • Radioactive medical devices
  • High-Risk, Class III devices
  • Excluded barrier contraceptives, including condoms
  • Devices, including medicinal and those of biological origin

Kingdom of Saudi Arabia

In the Kingdom of Saudi Arabia (KSA), you can begin the medical device registration process if your company has regulatory approval from one of the founding members of the Global Harmonization Task Force (GHTF). The five founding members are 1) the USA, requiring a 510(k) or PMA; 2) Canada, requiring a Medical Device License; 3) Europe, requiring CE Marking; 4) Australia, requiring Australian Register of Therapeutic Goods (ARTG); and 5) Japan, requiring Japanese Pharmaceutical Affairs Law (JPAL) certification or approval. The next step is to select an Authorized Representative in the KSA and submit a Medical Device Marketing Authorization (MDMA) application. For low and medium-risk device classifications (i.e., – Class I, IIa, and IIb), you may begin marketing your device in the KSA before obtaining formal regulatory approval (this regulation is subject to potential change) for higher-risk devices (i.e., – Class III), you must first obtain MDMA certification before distribution of the device in the KSA. The medical device regulations for the KSA are interim regulations. You can verify the current regulations by visiting the Saudi Food and Drug Authority (SFDA) website (http://bit.ly/SFDAMedicalDevices).

Other Countries

Many other countries have alternate, abbreviated processes similar to Australia and the KSA if your medical device is already approved by one of the GHTF countries. Often, this is stated as “country of origin approval.” Countries recognizing country of origin approval that offer an abbreviated approval process include Argentina, Singapore, China, etc. These countries do not merely “rubber stamp” the approval, but the approval process is less rigorous.

A CE certificate is insufficient if your product is manufactured in the US but you do not have a PMA or 510(k) issued by the US FDA. Your company must establish a country of origin status in Europe to take advantage of the abbreviated approval processes. This is sometimes done by establishing a European facility, but the CE certificate must be issued to the European facility. Other workarounds have been developed, but that is beyond the scope of this blog.

2 EU Candidate Member States with Competent Authorities

These two countries below are candidate member states for joining the EU. These countries are not signatories, but both countries have established a competent authority for reporting recalls and vigilance related to medical devices distributed within their borders. Turkey has also established four Notified Bodies.

  1. Iceland
  2. Turkey

3 EEA Signatories with Competent Authorities

For a long time, Switzerland was neither a member of the EU nor a signatory. However, in 2008, Switzerland became the 25th country to sign the Schengen Agreement, allowing people to pass between countries without border controls. All three countries below have established a competent authority. Switzerland has established five Notified Bodies, and Norway has two.

  1. Liechtenstein
  2. Norway
  3. Switzerland

28 EU Member States with Competent Authorities

The list below identifies the 28 members of the EU. The date in parenthesis is the year that each member joined the EU. These countries have competent authorities that regulate medical devices, and many of these countries have established Notified Bodies. Germany, Italy, and the UK have the most significant Notified Bodies.

  1. Austria (1995)
  2. Belgium (1952)
  3. Bulgaria (2007)
  4. Croatia (2013)
  5. Cyprus (2004)
  6. Czech Republic (2004)
  7. Denmark (1973)
  8. Estonia (2004)
  9. Finland (1995)
  10. France (1952)
  11. Germany (1952)
  12. Greece (1981)
  13. Hungary (2004)
  14. Ireland (1973)
  15. Italy (1952)
  16. Latvia (2004)
  17. Lithuania (2004)
  18. Luxembourg (1952)
  19. Malta (2004)
  20. Netherlands (1952)
  21. Poland (2004)
  22. Portugal (1986)
  23. Romania (2007)
  24. Slovakia (2004)
  25. Slovenia (2004)
  26. Spain (1986)
  27. Sweden (1995)
  28. United Kingdom (1973)

Which Countries Require CE Marking of Medical Devices? Read More »

What Does the CE Mark Mean, and What is its Purpose?

CE Marking Examples

The author answers the question of what does the CE Mark means, what its purpose is related to medical devices and regulatory requirements, if applicable.

Click here to subscribe to our blog feed

To facilitate trade throughout the European Economic Area (EEA), products need to be identified as compliant with regional and national regulations. In the EEA, this identification is the CE Mark. “CE” is not an acronym. The mark indicates compliance of your product with the essential requirements in the applicable directive. In the case of medical devices, there are three directives:

  1. Medical Device Directive, 93/42/EEC (http://bit.ly/M5MDD),
  2. Active Implantable Medical Devices Directive, 90/385/EEC (http://bit.ly/AIMDDirective)
  3. In Vitro Diagnostics Directive, 98/79/EC (http://bit.ly/currentIVDD).

Prior to the existence of these three directives, medical devices were compliant with the regulations of individual member states. These regulations were extremely detailed and created a barrier to the transport of products between the member states. With the implementation of the new approach directive (http://bit.ly/Resolution85), companies were able to CE Mark medical devices in accordance with one of the three device directives, and medical device products began to flow smoothly throughout the EEA.

Notified Body Numbers

The images at the top of this blog posting are examples of CE Marks from two of the largest medical device Notified Bodies. The four-digit numbers identify the Notified Body (NB) that issued the CE Certificate for the medical device. This number is only used for medical devices requiring NB involvement. Therefore, non-sterile Class I medical devices that do not have a measurement function are required only to have the “CE” on their labeling. All other medical devices are required to have the “CE” with the NB four-digit number. If one of the Competent Authorities (CAs), the equivalent to the U.S. FDA in each member state, wants to determine which Notified Body is authorizing the CE Marking of a medical device, the CA will look-up the four-digit number on the following NB database (http://bit.ly/NBDatabase).

How to Reproduce the Mark

It is the legal manufacturer’s responsibility to design their labeling with the CE and NB number—if applicable. This labeling is included in the company’s Technical File, and the NB reviews the Technical File for compliance with the essential requirements in one of the three device directives. For medical devices, the instructions for CE Marking are defined in Annex XII of 93/42/EEC. For active implantable devices, the requirements are found in Annex 9 of 90/385/EEC. For in vitro diagnostic devices, the requirements for CE Marking are found in Annex X of Directive 98/79/EC.

These three Annexes are identical and provide a graduated drawing showing the exact proportions of the “C” and “E” relative to one another. These Annexes also state that “”The various components of the CE marking must have substantially the same vertical dimension, which may not be less than 5 mm.”” You can obtain a free download of the mark on the Europa website (http://bit.ly/DownloadCE).

The four-digit NB number is intended to be the same boldness and font as the “”CE”” characters. Therefore, NBs have interpreted the requirement to specify numbers that are at least half the height of the “C” and “E”—or at least 2.5 mm. Each NB also provides instructions to legal manufacturers on how to present the CE characters with their four-digit NB number. Usually, there are a couple of different orientations that are allowed by the NB. For small products, it may not be possible to mark the device with a “C” and “E” that is at least 5 mm. Therefore, the directives waive this minimum dimension for small-scale devices. Most companies, however, will place a “C” and “E” on their labeling that is at least 5 mm in height, instead of marking parts with a “CE” that is illegible.

Use and Misuse of CE Marking

Most companies want to use CE Marking on all product labeling, even for products sold outside the EEA, because other countries recognize it and associate it with safety and performance. It is also acceptable to use the “CE” in product literature. However, it is important that it appears next to product images or descriptions that have a valid CE Certificate. It is not acceptable to use the “CE” in a way that it might imply that other products have a CE Certificate when the products do not. It is also not acceptable to use the “CE” in a way that it might imply a corporate entity is “CE Marked.” CE Certificates are for products—not for companies.

Please join us on Social Medialinkedin 1 Homegoogleplus Hometwitter Home

What Does the CE Mark Mean, and What is its Purpose? Read More »

CE Marking Routes to Regulatory Approval

ce marking routesReviewing the conformity assessment process contained in Annexes VII, II, V, and III related to seeking CE Marking regulatory approval.

Click here to subscribe to our blog feed

CE Marking of medical devices requires technical documentation (i.e., – a Technical File or Design Dossier)—regardless of the device classification. However, the classification of the device has a significant impact on the regulatory approval pathways available to your company. Therefore, the first step in the process of CE Marking of medical devices is to determine the risk classification (https://medicaldeviceacademy.com/ce-marking-faqs/). There are four device classifications: Class I, IIa, IIb, and III.

Once your company has determined the risk classification of the device, then you must decide which conformity assessment procedure you will follow to receive CE Marking approval. The conformity assessment procedures available for each classification are identified in Article 11 of the Medical Device Directive (http://bit.ly/M5MDD), and additional detail is provided in the various Annexes (i.e., – Annex II, III, IV, V, VI, and VII). The following table below summarizes the options for each classification:

If your product is a Class I device that is non-sterile and non-measuring, then you will not require a Notified Body (NB). However, all other products will require your company to select a NB (http://bit.ly/SelectingRegistrar).

Annex VII – Declaration of Conformity

If your company does not require NB involvement, then you can issue a Declaration of Conformity In accordance with Annex VII. You must also register your product with one of the Competent Authorities (CA) in Europe. CAs are the U.S. FDA equivalent in each EU member state. The following is a list of contact information for all the CAs: http://bit.ly/ContactPoints. If your company has no physical presence in Europe, you must also select a European Authorized Representative (AR). I recommend selecting an AR who is one of the 15 members of the European Association of Authorized Representatives (http://bit.ly/EAARMembers).

Annex II – Full Quality Assurance

Most companies use the Annex II conformity assessment process to achieve CE Marking. In this process, the NB reviews your Technical File for conformity and your quality system for conformity with regulatory requirements in the applicable directives. As part of the Annex II process, the NB will audit your design process to ensure that you have adequate design controls and that your process for establishing and maintaining a Technical File is sufficient. Once your company has adequately addressed any findings from the audit, the NB will issue your company a Full Quality Assurance (FQA) CE Certificate in accordance with Annex II.3. Once you have the Certificate, your company will be able to launch new products without prior approval from the NB. The only requirement is that the new products are within the scope of the Annex II.3 certificate.

Annex V – Production Quality Assurance

The Annex V conformity assessment process is the most common route to CE Marking for companies that outsource product design to a third party. If your company outsources design, Clause 7.3 is excluded from your ISO 13485 Certification, and you cannot demonstrate “Full Quality Assurance.” Therefore, the NB will issue an Annex V certificate for “Production Quality Assurance.” Annex IV and Annex VI are alternate conformity assessment procedures, but these are used less frequently for medical devices and are outside the scope of this blog.

Annex III – Type Examination

The Annex III conformity assessment process is a type of examination that is performed for higher-risk devices where the company does not have an Annex II certificate. This type of examination involves a review of your company’s design dossier, and the NB issues a Type Examination CE Certificate. This Certificate cannot be used alone for CE Marking. Type Examination Certificates must be used in conjunction with another CE Certificate, such as the Annex V certification for Production Quality Assurance. This combination would be used for Class IIb and Class III devices instead of an Annex II CE Certificate.

If your company needs help with CE Marking, including training on the medical device directive, please contact Medical Device Academy at rob@13485cert.com. We are also developing a webinar series for this purpose. If you are interested in more services, try viewing the following blog category page: CE Marking.

CE Marking Routes to Regulatory Approval Read More »

How to write a CE Classification Rationale

technical_fileThis article explains how to write a CE classification rationale for medical devices compliant with Annex VIII of the EU MDR.

Note: Blog page is under construction

 CE Marking of medical devices requires technical documentation (i.e., – a Technical File or Design Dossier). One of the requirements of this technical documentation is to establish the risk class of a device in accordance with the classification rules in Annex IX. The requirement to include this classification rationale in the Technical File is not well defined in Article 9, Classification, of the Medical Device Directive (http://bit.ly/M5MDD), but the guidance document for Technical Documentation (http://bit.ly/NBMED251Rec5) clearly defines this requirement in Section 3.2 (viii). Establishing a classification rationale is the first step to establishing the regulatory pathway that will be required to CE Mark your medical device.

Click here to subscribe to our blog feed

What are the CE Marking device classifications?

There are four different classifications of medical devices in Europe: Class I, Class IIa, Class IIb and Class III. These four classifications are also referred to as “risk class.” The lowest risk classification is Class I, and the highest risk classification is Class III. Class IIa is considered a “medium risk,” while Class IIb is considered a “high risk” medical device. Notified Body involvement and a CE Certificate from the Notified Body is required for almost all medical devices distributed in Europe, however, Class I devices that are non-sterile and do not perform a measurement function do not require Notified Body involvement. Class I devices that are sterile are often referred to as “Class I-S,”but this is not a term used in the Directive. The same is true of “Class I-M” for Class I devices with a measurement function.

Applying the classification rules for medical device CE Marking

In order to apply the classification rules as defined in Annex VIII of the MDR, the following questions must be answered for the device or device family:

  1. Is the device invasive? –  Invasiveness of a device is an important criterion, because non-invasive devices are generally Class I, and there is typically no Notified Body involvement required for these devices.
  2. Is the device surgically invasive, or invasive with respect to body orifices?  If a device is surgically invasive with respect to a body orifice, Rule 5 is the most likely classification rule. For all other devices that are surgically invasive, the duration of use is important
  3. How long is the device used inside the body? The primary difference between Rules 6, 7 and 8 is the duration of use. In general, permanent implants are subject to Rule 8, and are Class IIb devices. The other surgically invasive devices are generally Class IIa devices. Devices under Rule 6 are for “transient” use (i.e., – intended for continuous use for less than 60 minutes). Devices under Rule 7 are for “short-term” use (i.e., – intended for continuous use for between 60 minutes and 30 days.). There are multiple exceptions to each rule, and these exceptions should all be considered.
  4. Is the device electrically powered (i.e., – an active device)? Active devices are subject to rules 9, 10, 11 and 12.
  5. Do any of the “Special Rules” apply? It is recommended to actually start with Rules 13-18 to ensure that one of the special rules do not apply. For example, if you are making blood bags, there is no need to read anything in Annex VIII except Rule 18.

Guidance Document for Classification

Annex IX defines the classification rules for Europe, but there is also a guidance document (http://bit.ly/EUClassification) published that helps to explain the classification rules with examples. This guidance document is extremely important, because it provides clarification of rules based upon interpretations that have been made by Competent Authorities with Notified Bodies and companies. For example, critical anatomical locations are defined in Section 3.1.3: “For the purposes of the Directive 93/42/EEC, ‘central circulatory system’ means the following vessels:…”.

When you write a classification rationale for your technical documentation, it is important to reference this document—as well as Annex IX of the MDD. Your rationale should address each of the questions above for applying the classification rules. In addition, your rationale should indicate that none of the “Special Rules” (Rules 13-18) are applicable to your device or device family.

Mixed Classifications

It is possible to have a device family, contained within one Technical File or Design Dossier, that has more than one Classification. For example, you could choose to group a family of vascular grafts together in one Technical File that are permanent implants and non-absorbable. Normally, these devices would be Class IIb in accordance with Rule 8. However, if one or more of the grafts is intended for vessels included in the central circulatory system, then these grafts would be Class III devices. If a graft can be used for either indication, then the higher classification should be applied.

EU MDR – Regulation (EU) 2017/745

On September 26, 2012, the European Commission released a draft proposal for a new medical device regulation. The expected implementation transition period for these proposed regulations is 2015-2017. In Annex II of the proposed regulations (http://bit.ly/EUProposal), it specifies that the risk class and applicable classification rationale shall be documented in the technical documentation. This appears as item 1.1e) under the heading of “Device description and specification.”

The MDD defines the classification rules for medical devices in Annex IX, while in the proposed EU regulations classification, rules are now in Annex VIII. The MDD also has 18 rules, while the proposed regulations have 21 rules. In order to download a Gap Analysis of the Classification Rules for CE Marking, please visit the following page on this website: http://bit.ly/gapanalysiscmda.

If you need assistance with medical device CE Marking, or you are interested in training on CE Marking, please contact Medical Device Academy at: rob@13485cert.com. Medical Device Academy is developing a webinar series specifically for this purpose. You can also call Rob Packard by phone @ 802.258.1881. For other blogs on the topic of “CE Marking,”please view the following blog category page: http://robertpackard.wpengine.com/category/ce-marking/

How to write a CE Classification Rationale Read More »

Scroll to Top