Medical Device Shortage Reporting

The FDA and Health Canada both have executive-level orders requiring medical device shortage reporting or supply-chain disruptions.

In a previous article, we discussed supply-chain disruptions and mentioned that there might be medical device shortage reporting requirements if that disruption causes a market shortage of the manufactured device. Both the United States and Canada have reporting requirements for supply disruptions or the market’s ability to meet the demand of specific types of devices.

Click here to subscribe to our blog feed

Both the U.S. FDA and Health Canada have executive-level orders that require reporting of shortages or disruptions to the supply of medical devices deemed necessary for the COVID-19 Health Emergency. There is some overlap, but each country is monitoring and experiencing shortages and disruptions of different devices.

Where did medical device shortage reporting responsibilities come from?

Check 21 CFR 820, ISO 13485:2016, and even peek at SOR 98-282 and see if you can find your obligations for reporting. Go ahead. I’ll wait… Not much in there, right? Adverse events, complaints, etc., but not market shortages.
Medical device shortage reporting is specific to health emergencies. The U.S. FDA and Health Canada happen to be two authorities having jurisdiction with reporting requirements for shortages concerning the COVID-19 Health Emergency. However, there may be others, so having your organization’s regulatory affairs manager verify the reporting requirements for the markets in which you are engaged might not be bad.

U.S. FDA 506J reporting-

fda u.s. food and drug adminstration logo - marks the begining of the section specific to US FDA requirements
U.S. FDA logo


In the United States, an Amendment to the U.S. Food, Drug, and Cosmetics Act requires regulatory reporting by medical device manufacturers to the U.S. FDA. It is sometimes called 506J reporting for the Section of the U.S. FD&C Act where it is located.

You will find the statutory requirements outlined within 21 USC 356J.

Screen capture of where medical device shortage reporting statutory requirements are outlined (i.e. 21 USC 346J).
21 USC 356J Discontinuance or interruption in the production of medical devices

For the full text read, 21 USC 356j: Discontinuance or interruption in the production of medical devices. (Interestingly enough, the website where this information is available is not an HTTPS site, so visit at your own discretion).

http://uscode.house.gov/browse.xhtml

What devices are subject to 506J reporting?

There are two types of devices that the FDA is monitoring. “Critical” devices and an FDA-published list of devices for which COVID-19 is causing a higher than expected demand.

The FDA has released a guidance document that contains criteria for what is considered to be a “Critical Device”. This includes devices such as those used during surgery, emergency medical care, and those intended to treat, diagnose, prevent, or mitigate COVID-19.

screenshot of the critical device criteria for 506J medical device shortage reporting to the FDA
Screenshot of the Critical Device Criteria for 506J reporting

There is also a published list of concerned devices that the FDA is specifically monitoring. The FDA website lists these devices by product code, but include the following device types;

  • Clinical Chemistry Products
  • Dialysis-Related Products
  • General ICU/Hospital Products
  • Hematology Products
  • Infusion Pumps and Related Accessories
  • Microbiology Products
  • Needles and Syringes
  • Personal Protective Equipment (PPE)
  • Sterilization Products
  • Testing Supplies and Equipment
  • Ventilation-Related Products
  • Vital Sign Monitoring
a screenshot of some of the devices listed in the FDA 506J medical device shortage reporting list
Screenshot of the FDA Shortage List

Understandably this process may not be intuitive, and for this, the FDA has released a guidance document that addresses;

  • Who must make the notification
  • When you should make a notification
  • What information needs to be included within your 506J notification
  • How to make a notification, and
  • Penalties for failure to make a notification

The referenced product codes may not be an all-inclusive list or entirely up to date. The best suggestion for full compliance is to go straight to the source of the regulation, in part because noncompliance can result in enforcement action from the FDA. If you think that your device might require notification to the FDA but isn’t in the reference table, you should contact the FDA for notification clarification. Below is the quote from the FDA website, and it includes the contact email for asking these specific questions to ‘the agency.’

“If a device type is not included in this table, but you believe it requires a notification under section 506J of the FD&C Act, or if you have questions regarding the device types in this table, you should contact FDA at CDRHManufacturerShortage@fda.hhs.gov and include “Question” in the subject line of the email.”

Link to the FDA Guidance Document for 506J Reporting- HERE

How to make a 506J report to the U.S. FDA?

The FDA accepts 506J reports in multiple ways. For example, you may use the 506J Reporting web form or submit a notification by email directly to (Include Email Here). In addition, Medical Device Academy has developed a Work Instruction and Form to determine if your company is experiencing a reportable discontinuance or meaningful disruption in manufacturing a medical device as well as compiling the report for submission.

There are a few methods of notification, a web form for individual notifications and spreadsheet options for multiple notifications at once, or emailing a report directly to the FDA reporting email included below;

CDRHManufacturerShortage@fda.hhs.gov

Screenshot of the FDA 506J reporting webforms for medical device shortage reporting
Screenshot of the FDA 506J reporting Webforms from https://fdaprod.force.com/shortages

It is for this process that Medical Device Academy developed WI-010 506J Shortage Reporting to the U.S. FDA. This work instruction and associated form, FRM-053 506J Reporting Form are designed to walk you through the process of determining reportability and compiling the information necessary to either complete the webform or email the report directly to the shortage reporting email.

Medical Device Shortage Reporting to Health Canada

health canada logo w french - section of the article specific to Canada
Health Canada logo

Rather than discontinuance and disruption of manufacture, Health Canada is monitoring for shortages of specific devices. Therefore, Health Canada wants Medical Device Shortage Reports regardless of the reason for the shortage. It also shows that this is not identical reporting of the same conditions to two different authorities. Health Canada will also accept reports from Importers because the frame of reference is Canada’s supply of medical devices concerning Canada’s needs.

As an Authority Having Jurisdiction, Health Canada also has reporting requirements for medical device shortage reporting of specific types of medical devices. Health Canada is also an independent authority that uses a different device classification system than the U.S. FDA.

The table below shows the device types by their classification level that HC requires supply chain disruption notifications for. This information is current as of September 5th, 2021, and the link below will take you to the HC website page for the most up-to-date list.

https://www.canada.ca/en/health-canada/services/drugs-health-products/medical-devices/shortages/covid19-mandatory-reporting.html

Class I Medical Devices
Masks (surgical, procedure or medical masks) – Level 1, 2, 3 (ATSM)
N95 respirators for medical use
KN95 respirators for medical use
Face shields
Gowns (isolation or surgical gowns) – Level 2, 3 and 4
Gowns (chemotherapy gowns)
Class II Medical Devices
Ventilators (including bi-level positive airway pressure or BiPAP machines, and continuous positive airway pressure or CPAP machines)
Infrared thermometers
Digital thermometers
Oxygen Concentrators
Pulse Oximeters (single measurement)
Aspirators/suction pumps (portable and stationary)
Laryngoscopes
Endotracheal tubes
Manual resuscitation bags (individually or part of a kit)
Medical Gloves – Examination and Surgical (Nitrile, Vinyl)
Oxygen Delivery Devices
Class III Medical Devices
Ventilators (including bi-level positive airway pressure or BiPAP machines)
Pulse Oximeters (continuous monitoring)
Vital Signs Monitors
Dialyzers
Infusion Pumps
Anesthesia Delivery Devices
Class IV Medical Devices
Extracorporeal Membrane Oxygenation (ECMO) Devices
List of ‘Specified Devices’ that Health Canada is monitoring for shortage reporting

One of the things that Health Canada does an excellent job of is defining its expectations. In the Second Interim Order Respecting Drugs, Medical Devices and Foods for a Special Dietary Purpose in Relation to COVID-19, it is explained the Manufacturers or Importers should report to the Minister actual or expected shortages of the device, OR components, accessories, or parts. These notifications must be made within 5-days of becoming aware of the shortage or the anticipated shortage date. Update reports must be made within 2-days of becoming aware of new information regarding the shortage, and a closing report must be made within 2-days of the end of the shortage.

(This link is to the HC website for the 2nd Interim Order referenced above)

https://www.canada.ca/en/health-canada/services/drugs-health-products/covid19-industry/drug-medical-device-food-shortages/interim-order-2021.html

How to make a shortage report to Health Canada?

These reports are submitted online through the Health Canada Website. They have an entire section dedicated to medical device shortages, and the reporting links can be found there (Link here). If you have any questions or are on the fence about notification, you can email Health Canada at MD.shortages-penurie.de.IM@canada.ca.

Where you can find Health Canada's website links for manufacturers to enter medical device shortage reporting or to report the end of a shortage.
Health Canada Webforms for reporting a shortage and the end of a shortage

The webform for reporting a shortage is the same webform that is used for providing update reports to Health Canada as well. This is both for manufacturers of specified medical devices as well as importers.

1 thought on “Medical Device Shortage Reporting”