What is the Scope of IEC 60601-1:2005 (3rd edition)?

60601 300x274 What is the Scope of IEC 60601 1:2005 (3rd edition)?This blog will help you determine if and how the IEC 60601-1 Standard applies to your medical electrical product.  Definitions, third-party testing, etc. discussed.

One of the first questions clients ask before a project starts is, “Does this medical electrical product fall under IEC 60601-1?” Another common question clients ask is whether IEC 60601-1 applies to battery operated medical devices. To answer these questions, we must review and understand the scope (sub-clause 1.1) of IEC 60601-1 to determine if and how the IEC 60601-1 Standard applies to a medical electrical product.

The title of IEC 60601-1:2005 (3rd edition) ( is Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. The IEC 60601-1 Standard itself states, “This…Standard applies to…MEDICAL ELECTRICAL EQUIPMENT and MEDICAL ELECTRICAL SYSTEMS…referred to as ME EQUIPMENT and ME SYSTEMS.”

*Note: ALL CAPITAL LETTERS identifies a defined term for the IEC 60601 series of standards within this blog. “IEC” is an acronym for the International Electrotechnical Commission ( IEC is a non-profit, non-governmental international standards organization that prepares and publishes International Standards for all electrical, electronic and related technologies.

All clause references in this blog are to both IEC 60601-1:2005 (3rd edition) and IEC 60601-1:2005 (3rd edition) + Amendment 1:2012, or the consolidated version IEC 60601-1:2012 (edition 3.1), but the actual text comes from edition 3.1.


The ME EQUIPMENT definition (sub-clause 3.63) is:


a)    provided with not more than one connection to a particular SUPPLY MAINS; and

b)    intended by its MANUFACTURER to be used:

1)    in the diagnosis, treatment or monitoring of a PATIENT; or

2)    for compensation or alleviation of disease, injury or disability.”

and “…having an APPLIED PART or transferring energy to or from the PATIENT or detecting such energy transfer to or from the PATIENT…”.

Note: The above definition aligns well with the definition for medical electrical devices in the European Medical Device Directive (

From the definition above, we know that a device can have up to one power cord, or be hard wired to the building’s power by one power line, and/or be battery powered.

We also need to understand the term APPLIED PARTS (sub-clause 3.8): “Part of…ME EQUIPMENT…in NORMAL USE necessarily comes into physical contact with the PATIENT for…” the device “…to perform its function.” Classification of the different types of APPLIED PARTS and other classifications (there are eight different classification criteria in the IEC 60601 Standard) need to be completed early in the process of setting up a test plan for design verification to determine the applicable testing requirements, and to develop an appropriate test plan.

Examples within Scope of IEC 60601-1:2005

Examples of electrical medical products fitting the definition above are broad and include battery-operated thermometers, gamma imaging systems, endoscopic cameras and infusion pumps. IEC 60601-1 may also apply to many ACCESSORIES (sub-clause 3.3) used with ME EQUIPMENT.

Examples not within Scope of IEC 60601-1:2005

The scope of IEC 60601-1 also identifies which devices are not included in the IEC 60601 series:

  • “in vitro-diagnostic equipment…IEC 61010 series;
  • implantable parts of active implantable(s)…ISO 14708 series…;
  • medical gas pipeline systems…ISO 7396-1…”
IEC 60601 & ISO 13485

Design verification must confirm that design outputs (i.e., – product specifications) meet design inputs (e.g., – product must meet IEC 60601-1 requirements).

Many projects don’t identify all the applicable IEC 60601 standards. This could cause multiple nonconformances during an audit by the national regulatory body (i.e., FDA, EU Notified Body), or that you don’t obtain approval to sell and distribute your device from the national regulatory body.

A test plan, with multiple test protocols, is developed from the product specification. This test plan should identify all the requirements of the IEC 60601 series of standards, in addition to any other applicable standards and regulations that apply to the device prior to performing device testing.

Third-Party Testing

IEC testing can be performed by a third-party test house (i.e., – a safety certification agency, such as a UL, CSA, TÜV SÜD),  an independent test lab (i.e., – Global EMC Labs, O’Brien Compliance Management, Medical Equipment Compliance Association), or the manufacturer can conduct the testing if they have proper equipment, trained personnel and a good understanding (i.e., – used the Standard on several projects, and successfully tested previous similar electrical medical device(s) by a third-party test lab, and been approved by a national regulatory body) of the standard(s).

Design verification reports generated from the test process are either the applicable IEC 60601 and IEC 80601 series of standards test report forms ( and, or the manufacturer’s generated test reports.

If your company needs help with IEC 60601-1 gap analysis, preparation of the risk management file for the third edition, or training on the Standard, please contact Leo Eisner at: We are also developing a webinar series on IEC 60601-1, 3rd edition.


Please Join Medical Device Academy on Social Medialinkedin 1 What is the Scope of IEC 60601 1:2005 (3rd edition)?googleplus What is the Scope of IEC 60601 1:2005 (3rd edition)?twitter What is the Scope of IEC 60601 1:2005 (3rd edition)?

Posted in: IEC 60601

Leave a Comment (10) ↓


  1. Hossein Khosropour January 14, 2014

    Do video endoscopes need an IEC certificate? Considering that video endoscopes do in vitro-diagnostic and there is not any transfer of energy to or from the PATIENT.

    Thank you for your reply.

    • Rob Packard January 16, 2014

      Great question, but my friend +Leo Eisner is the expert on this. I believe that electrical safety requirements would apply, because the CCD in the endoscope is electrically powered. However, there are many classifications of electrical medical equipment. I checked the FDA classification and IEC 60601-1 was not listed, but I will ask Leo to add a comment here.

  2. Augustus Shanahan July 31, 2015

    From what I understand, Endoscopes are included in IEC 60601-1, and in addition they have their own Particular Standard IEC-60601-2-18.

    • Rob Packard November 6, 2015

      Most products have multiple IEC 60601 standards that apply. The IEC 60601-2-18 is a particular standard. Each product category could fall under one or more of these particular standards.

  3. Jamie July 1, 2016

    Does an electrical surgical instrument system used for fat harvesting and body contouring fall within the scope of IEC 60601-1?

    • Rob Packard November 4, 2016

      Low level lasers for aesthetic purposes and lipoplasty systems are both regulated as Class II medical devices that require electrical safety testing under IEC 60601-1. Sometimes the collateral standards are not specified under the product classification codes, but you will find that there is electrical safety testing listed in the 510(k) summary of most of the predicate devices.

  4. Edenia Guevara May 25, 2017

    How do I know which clauses, sub-clauses of the standard are applicable to my Medical Device? Battery Powered, hand held device, for Hospital Environment.

    • Rob Packard August 17, 2017

      You have to review the standard itself and you need specific information about your device. For battery powered devices, only certain clauses are applicable, but this is stated in the standard. Another factor is whether the device has an applied part, and finally the use environment can change the applicable standards. For example, lab equipment falls under a different standard, but labs are still within a hospital.

  5. Douglas Harvey August 18, 2017

    What are the requirements for a device to be considered mains powered? Is a battery powered camera considered mains powered if it is being charged during operation? Would an isolating switch that is enabled during use be enough to call it a battery powered device?

    • Rob Packard September 24, 2017

      This response was provided by Leo Eisner of Eisner Safety.

      When the device is disconnected from the power source, and not charging, it will fall under INTERNAL POWER (definition 3.45 – see definition below.) When the device is connected to the MAINS POWER, it will fall under mains powered. If you refer to 4.10.1 below, you will see you can connect to SUPPLY MAINS, connection to a separate power supply or be powered by an INTERNAL ELECTRICAL POWER SOURCE. Or you can combine in to a combination of these sources.
      Electrical power source for operating equipment that is a part of the equipment, and which produces electrical current from some other form of energy. Examples include chemical, mechanical, solar or nuclear.
      4.10.1 Source of power for ME EQUIPMENT
      ME EQUIPMENT shall be suitable for connection to a SUPPLY MAINS, be specified for connection to a
      separate power supply or be powered by an INTERNAL ELECTRICAL POWER SOURCE. Alternatively, a combination of these sources may be used.

      The more specific questions asked required additional input that was communicated by Leo directly to Douglas Harvey.


Leave a Comment

Time limit is exhausted. Please reload the CAPTCHA.


Get every new post on this blog delivered to your Inbox.

Join other followers:

Simple Share Buttons
Simple Share Buttons