Essential Requirements

What are the Essential Requirements for Medical Device CE Marking?

ER Table1 What are the Essential Requirements for Medical Device CE Marking?

The author reviews the essential requirements for medical device CE marking. Common mistakes to avoid, and the proposed EU regulations are also discussed.

Essential Requirements (ERs) are the requirements for safety and performance specified in Annex I of the three medical device directives. ERs are divided into Part I (i.e., – general requirements) and Part II (i.e., – requirements for design and construction). Evidence of conformity must be provided for all general requirements in Part 1 for all devices—regardless of risk classification, design, or construction. The Design and construction requirements in Part 2 may be not applicable, depending upon your device.

When a Notified Body reviews your Technical File or Design Dossier for CE Marking, the auditor must verify that you have addressed each ER. This is typically demonstrated by providing an ER Checklist (ERC). You can find a template for an ERC on the International Medical Device Regulators Forum (IMDRF) website (http://bit.ly/IMDRFDoc) in Appendix A (see example in Figure 1 below) of the GHTF Guidance document explaining the use of a Summary Technical Document (STED) to demonstrate conformity with the principles of safety and performance (http://bit.ly/GHTFSTEDGuidance).

Figure 1 ERC Example What are the Essential Requirements for Medical Device CE Marking?

Figure 1: Example of an ERC

To demonstrate compliance with the ERs, you must provide the following information by filling in the four columns of the ERC:

  1. Applicability to your device,
  2. The method used to demonstrate conformity with the ER,
  3. Reference to the method(s) used, and
  4. Reference to the supporting controlled documents.
Subparts & Common Mistakes

Completing the ERC would be easy if there were only 13 ERs, but eight of the 13 ERs have multiple requirements. For example, ER 13.3 has 14 subparts (i.e., – 13.3a through 13.3n). Each subpart must be addressed when you complete the columns of the ERC table. If any of the parts in ER 7-13 do not apply to your device, you need to provide a justification. For example, ER 11 and its subparts do not apply to devices that do not emit radiation. This justification must be documented in the ERC for each subpart.

When you write your justification for the non-applicability of an ER, you need to be careful to provide a justification for each part of the requirement. For example, there are three sub-parts to ER 7.5. Each part is a separate paragraph, but these are not identified by a letter, as is done in ER 13.3 and 13.6. Instead, each subpart is a separate paragraph. Within those paragraphs, there is further room for confusion. For example, the third paragraph states that if you use Phthalates in a product that is intended for women or children, then you must provide a justification for its use in the technical documentation, in the instructions for use, within information on the residual risks for these patient groups (i.e., –women and children) and, if applicable, on appropriate precautionary measures.

Proposed EU Regulations

The proposed EU Medical Device Regulations (MDR) are organized into Articles and Annexes–just like the current EU Directives, and the ERs will still be the first Annex of the MDR. However, there will be 19 ERs instead of 13. The early reviews of the proposed regulations indicated that there were no significant changes. Still, I have learned the hard way that you should always go to the source and verify the information for yourself. The general organization of the Essential Requirements is still the same. Nevertheless, several significant changes will require providing additional documentation in your Technical File or Design Dossier for CE Marking. Most companies will probably submit a revised ERC to address the new requirements, but you may want to read Medical Device Academy’s review of the new ERs (http://bit.ly/NewERCGap) and prepare accordingly.

Essential Principles Checklists

Health Canada has an Essential Principles checklist (EPC) that is similar to the European ERC, and Australia has a similar document (http://bit.ly/EPCTGA) with only a few minor differences. The Global Harmonized Task Force (GHTF) created an earlier version in 2005 (http://bit.ly/EPSafetyPerf). Health Canada will typically accept your ERC developed for the European Medical Device Directive (MDD), but a gap analysis should be performed against the Australian Regulations.

Now that the ENVI vote has passed (http://bit.ly/ENVIVotepasses), I asked a new consultant working for me to create a template for the new Essential Requirements in the new EU MDR regulations. You can download the MDD ERC Template and the new EU MDR Template. This new template also indicates the items that were recently modified (see the red lines).

What are the Essential Requirements for Medical Device CE Marking? Read More »

EU Medical Device Directive: 6 New Essential Requirements

%name EU Medical Device Directive: 6 New Essential RequirementsEssential Requirements (ER) changes in the proposed EU Medical Device Regulations versus the ER in Annex I of the EU Medical Device Directive are reviewed.

Click HERE if you want to receive future “European Regulatory Updates” by email. Just provide your name, company, phone number, and email address where you would like to receive updates.

Annex I of the European Medical Device Directive (http://bit.ly/M5MDD) is titled “Essential Requirements.” Most companies demonstrate that their device meets the 13 Essential Requirements (ERs) by creating an Essential Requirements Checklist (ERC). I have no idea what the origin of the ERC is, but you know that regulators love tables and checklists. This particular checklist is so commonly used that the Global Harmonization Task Force (GHTF) included an example of an ERC, called an “Essential Principles Checklist” (EPC) at the end of a guidance document on how to create Summary Technical Documentation (STED) for In Vitro Diagnostic devices (http://bit.ly/STEDIVD)—which is now maintained on the IMDRF.org website.

On September 26, 2012, the European Commission released a proposal for new EU Medical Device Regulations (http://bit.ly/EUProposal). This proposal still includes ERs in Annex I, but there are 19 ERs in the proposal. One regulatory professional recently sent me a follow-up question in response to an audio seminar I conducted in November (). Her question was, “What are the six new ERs?”

A few of the early reviews of the proposal indicated that there were no significant changes. Still, I have learned the hard way that you should always go to the source and verify the information for yourself (i.e., – Genchi Genbutsu). Here’s what I found:

General Requirements (ER 1-6a)

  1. No real change to this requirement.
  2. This requirement was reworded to clarify the intent (see Annex ZA of EN 14971:2012 for more info @ http://bit.ly/ISO14971-2012changes).
  3. It appears as though the Commission thought the current ER 3 was redundant, and the requirement was addressed by ER 1 and ER 5 already.
  4. This is now the new ER 3, and the requirement now clarifies how Notified Bodies shall apply this requirement in cases where a lifetime of the device is not stated.
  5. This is now the new ER 4, and there is no real change.
  6. This is now the new ER 5, and the wording has been clarified.

ER6a is conspicuously missing from the proposed ERs, but don’t get excited. Clinical evaluations are still required as part of the Technical Documentation in Annex II, Section 6.1c: “the report on the clinical evaluation in accordance with Article 49(5) and Part A of Annex XIII.”

Chemical, Physical & Biological Properties (ER 7)

ER 7.1 has one new requirement: “d) the choice of materials used, reflecting, where appropriate, matters such as hardness, wear and fatigue strength.” ER 7.2 and 7.3 remain unchanged. ER 7.4 has been simplified to what is proposed as the new, shorter ER 9. ER 7.5 is now the new ER 7.4, and the changes reflect the current status of phthalate regulations and similar issues. ER 7.6 is now the new ER 7.5, but there is no change to the content. The new ER 7.6 requires that manufacturers address risks associated with the size and properties of particles, especially nanomaterials. The changes related to this section will impact certain device types more than others—such as orthopedic implants.

Infection & Microbial Contamination (ER 8)

ER 8 is still ER 8, but ER 8.1 is now prescriptive regarding design solutions, and the current ER 8.2 is now the new ER 10. The new ER 10 is expanded and references the new EU regulations regarding devices manufactured utilizing tissues or cells of animal origin: Commission Regulation (EU) No 722/2012 of August 8, 2012 (http://bit.ly/AnimalTissueReg). The new ER 8.2 is a new requirement that was an oversight of the MDD, and the new ER 8.7 now clarifies that the labeling must differentiate sterile and non-sterile versions of the product; packaging is no longer an acceptable mechanism for differentiation. The balance of ER 8 remains unchanged.

Construction & Environmental Properties (ER 9)

This ER is now identified as the new ER 11, and this section is expanded. This reflects the emphasis on the need to evaluate the safety of devices with accessories, compatibility with other devices, and the effects of the use environment.

Devices with a Measuring Function (ER 10)

This ER is now identified as the new ER 12, but ER 10.2 from the current Directive appears to be missing. What’s up?

Take a look at the new ER 11. ER 10.2 is now the new ER 11.6.

Protection Against Radiation (ER 11)

This ER is now identified as the new ER 13, but there is nothing new.

Requirements for Devices Connected to or Equipped with an Energy Source (ER 12)

ER 12.1 and 12.1a are now ER 14. This section is specific to software requirements and has more detail than the current Directive. IEC 62304:2006, “Medical device software – Software life cycle processes,” is the Standard that will be expected by Notified Bodies as a reference for ER 14. ER 12.2 through ER 12.6 is now ER 15, but there is nothing new. Section ER 12.7 and its sub-parts are now addressed by ER 16. ER 12.8 and its subparts are now addressed by ER 17.

Information Supplied by the Manufacturer (ER 13)

This is now identified as ER 19: “Label and Instructions for Use.” This section is simplified from ER 13 (i.e., – there are fewer sections), but this ER does not seem to be any shorter. ER 19.1 has subparts a-g, and this ER section incorporates the concepts previously addressed by ER 13.1, 13.2, 13.4, and 13.5. ER 19.2 is a new and improved version of the previous ER 13.3 specific to labeling requirements. This labeling section is expanded from subparts “a” through “n” to “a” through “q.” The UDI requirement is subpart “h.” ER 13.6 is now ER 19.3 specific to the Instructions For Use (IFU). This section is expanded from subparts “a” through “q” to “a” through “t.”

The number of subparts to ER 19.3 doesn’t reflect the additional requirements for IFUs that are proposed by the Commission. The subsections of this part warrant special attention. Items that frequently are found missing from IFUs on the market today include:

  1. ER 19.3c – performance intended by the manufacturer
  2. ER 19.3h – installation and calibration instructions
  3. ER 19.3k – how to determine if a reusable device should be repaired/replaced
  4. ER 19.3m – restrictions on combinations with other devices
  5. ER 19.3o – detailed warning information
  6. ER 19.3p – information about safe disposal of the device
  7. ER 19.3t – notice to user/patient to report adverse events

ER 18 – Use by Lay Persons

This is a short section, but the requirement is new. There are no additional requirements for products intended for use by a layperson. The risk management report, design validation, and clinical evaluation report will need to include specific evidence to demonstrate conformity with this ER. The post-market surveillance plan for these products should carefully verify the accuracy of risk estimates. Post-Market Clinical Follow-up (PMCF) studies would be challenging in the past. Still, the prevalence of social media and product registration databases may facilitate conducting PMCF studies for these products in the future.

Australia & Canada

There is also an EPC that is required by the Therapeutic Goods Administration (TGA) in Australia, (http://bit.ly/EPCTGA) and  Therapeutics Product Directorate (TPD) in Canada (http://bit.ly/CanadianSTED). If you would like to learn more about the Essential Principles of Safety and Performance, you should also review the GHTF guidance document on this topic (http://bit.ly/EPSafetyPerf) on the IMDRF.org website. This 2012 version of the document supersedes GHTF/SG1/N041:2005.

I have observed the approval of products where the European ERC was submitted in place of an EPC for Australia and Canada. I guess they are a little more rational than some other regulators, but if you have experienced any “push back” regarding this approach, please share this by posting a comment or by sending an email.

If you need assistance with medical device CE Marking, or you are interested in training on CE Marking, please contact Medical Device Academy at rob@13485cert.com. Medical Device Academy is developing a webinar series specifically for this purpose. You can also call me by phone @ +1.802.258.1881. For other blogs on the topic of “CE Marking,” please view the following blog category page: http://robertpackard.wpengine.com/category/ce-marking/.

EU Medical Device Directive: 6 New Essential Requirements Read More »