How Declaration of Conformity (DoC) templates are created

This article about declaration of conformity for medical device CE Marking identifies three possible sources for creating a template and includes recommendations for making the transition to the proposed EU Medical Device Regulation (EMDR).

%name How Declaration of Conformity (DoC) templates are createdCurrent declaration of conformity requirements

Each of the three “New Approach” Device Directives require that manufacturers create a Declaration of Conformity (DoC). For the Medical Device Directive (MDD), this requirement is found in Annex II, IV, V, VI and VII. In fact, “Declaration of Conformity” is part of the title for each of those Annexes. A DoC is required for each CE Marked product family, but the MDD does not provide guidance for the format or content. Manufacturers need a controlled template.

Three sources of declaration of conformity examples

There are three possible sources for creating a declaration of conformity template:

  1. Copy another company’s declaration of conformity
  2. Authorized Representatives
  3. Annex III of the proposed European Medical Device Regulations

Most companies are using option number 1—often without knowing it. The problem with this approach is you have no regulatory reference to justify the basis for the content and format of your Declaration.

The second option is to ask your Authorized Representative (AR). The AR is supposed to verify that you have a DoC for each product family, and that the company has the required technical documentation for each product family. The AR must provide a DoC for each product they represent to the Competent Authority upon request. Therefore, some ARs provide manufacturers with a template for the DoC to ensure that your DoC meets requirements. This approach gives you a reference of external origin, and you can be sure that the template will meet current requirements.

My new recommendation for Creating A declaration of conformity

Option 3 is my new recommendation to companies. The proposed European Medical Device Regulation (EMDR) was released on September 26, 2012. The proposal provides a prescriptive template for the content of a DoC. This will eventually be required for every DoC. Annex III is only one page long and there are only 10 requirements:

  1. Name of the Manufacturer and the AR
  2. Declaration that DoC Issued under Sole Responsibility of Manufacturer
  3. Unique Device Identifier (UDI)
  4. Product Name and Catalog Numbers
  5. Risk Classification
  6. Declaration Statement of Conformity
  7. Reference to Harmonized Standards and Common Technical Specifications (CTS)
  8. Notified Body Name and Number, Description of Conformity Assessment Procedure and Certificate
  9. Additional Information
  10. Place and Date of Issue, along with the Name and Function of the Person Signing

Instead of waiting for the final approval of the EMDR, I recommend creating a declaration of conformity (DoC) template now that matches the proposed Annex III (you might want to add this to recommendations for improvement in your next management review).

Brigid Glass is the subject matter expert behind most of the conventions that our consulting firm uses for document templates. As with all of our forms, we assign a document control number (e.g., FRM-001) and the revision is “D1” for the first draft. The template is in table format in order to facilitate easier review by auditors and your AR. The left column of the table lists each of the required elements, and the right column is where the variable data is entered. Variable fields that are examples are in green font, and the content that does not require changing is in a black font. There are some comments to explain how to fill in the variable content, and several items that will change when the proposed regulation is approved are identified. The following harmonized symbols from ISO 15233-1 are used in the template, as well:

DoC How Declaration of Conformity (DoC) templates are created

Please click here to download Medical Device Academy’s declaration of conformity (DoC) template.

 

 

 

 

Posted in: CE Marking

Leave a Comment (3) ↓

3 Comments

  1. Jonathan Bretz January 15, 2014

    Robert,

    Excellent advise as always.

    Jon Bretz

    reply
    • Rob Packard January 16, 2014

      Thank you for the kind words of praise Jon. Call me sometime…802.258.1881.

      reply
  2. Kate Gonzalez May 25, 2018

    Very helpful. Thank you!

    reply